01 August 2007

Evaluation of analytic and motion-resistant performance of the Mindray 9006 Pulse Oximeter

Liming Peng, Cunliang Yan, Hong Lu, Yong Xia

Med Sci Monit 2007; 13(8): MT19-27 :: ID: 491814

Abstract

Background: The Mindray 9006 is a newly developed pulse oximeter. The performance of noninvasive monitoring of arterial oxygen saturation by pulse oximetry (SpO[sub]2[/sub]) and pulse rate (PR) with the Mindray 9006 was evaluated.
Material/Methods: Twenty healthy volunteers participated in this study. The Mindray 9006 was evaluated according to guidelines including HS3-A, EP5-A, and EP9-A2 published by the CLSI, and ISO 9919. A blinded side-by-side comparison of the Mindray 9006 and the Masimo MS-3 oximeter was also performed.
Results:The results demonstrated that the coefficients of variation (CV) of imprecision were less than 1% for SpO[sub]2[/sub] and less than 5% for PR analyzed by the Mindray 9006. Comparisons of methods showed there was no significant difference in SpO[sub]2[/sub]/PR analysis between the Mindray 9006 with Rapidlab 855 CO-oxymeter or 812 Electrocardiogram (ECG) monitor (p>0.05) and between the Mindray 9006 and Masimo MS-3 (p>0.05). There was good correlation for SpO[sub]2[/sub] between the Mindray 9006 and CO-oxymetry or the Masimo MS-3 (r>0.96) and for PR between the Mindray 9006 and ECG or the Masimo MS-3 (r>0.98). The accuracy as the root-mean-square (A[sub]rms[/sub]) was less than 3% for both SpO[sub]2[/sub] and PR between the Mindray 9006 and CO-oxymetry or between the Mindray 9006 and ECG. Also, there was no significant difference for SpO[sub]2[/sub]/PR between the motion and rest conditions (p>0.05).
Conclusions:The overall performance of the Mindray 9006 for SpO[sub]2[/sub] and PR analysis is excellent and compares favorably with the Masimo MS-3.

Keywords: Adolescent, Oxygen - metabolism, Oximetry - methods, Movement, Models, Statistical, Heart Rate, Evaluation Studies as Topic, Equipment Design, Blood Gas Analysis, Blood Gas Analysis, Oxygen - metabolism, Oximetry - methods, Movement, Models, Statistical, Heart Rate, Evaluation Studies as Topic, Equipment Design, Adolescent

Add Comment 0 Comments

Editorial

01 April 2024 : Editorial  

Editorial: Forty Years of Waiting for Prevention and Cure of HIV Infection – Ongoing Challenges and Hopes for Vaccine Development and Overcoming Antiretroviral Drug Resistance

Dinah V. Parums

DOI: 10.12659/MSM.944600

Med Sci Monit 2024; 30:e944600

0:00

In Press

06 Mar 2024 : Clinical Research  

Comparison of Outcomes between Single-Level and Double-Level Corpectomy in Thoracolumbar Reconstruction: A ...

Med Sci Monit In Press; DOI: 10.12659/MSM.943797  

21 Mar 2024 : Meta-Analysis  

Economic Evaluation of COVID-19 Screening Tests and Surveillance Strategies in Low-Income, Middle-Income, a...

Med Sci Monit In Press; DOI: 10.12659/MSM.943863  

10 Apr 2024 : Clinical Research  

Predicting Acute Cardiovascular Complications in COVID-19: Insights from a Specialized Cardiac Referral Dep...

Med Sci Monit In Press; DOI: 10.12659/MSM.942612  

06 Mar 2024 : Clinical Research  

Enhanced Surgical Outcomes of Popliteal Cyst Excision: A Retrospective Study Comparing Arthroscopic Debride...

Med Sci Monit In Press; DOI: 10.12659/MSM.941102  

Most Viewed Current Articles

17 Jan 2024 : Review article  

Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron Variant

DOI :10.12659/MSM.942799

Med Sci Monit 2024; 30:e942799

0:00

14 Dec 2022 : Clinical Research  

Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase Levels

DOI :10.12659/MSM.937990

Med Sci Monit 2022; 28:e937990

0:00

16 May 2023 : Clinical Research  

Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...

DOI :10.12659/MSM.940387

Med Sci Monit 2023; 29:e940387

0:00

01 Jan 2022 : Editorial  

Editorial: Current Status of Oral Antiviral Drug Treatments for SARS-CoV-2 Infection in Non-Hospitalized Pa...

DOI :10.12659/MSM.935952

Med Sci Monit 2022; 28:e935952

0:00

Your Privacy

We use cookies to ensure the functionality of our website, to personalize content and advertising, to provide social media features, and to analyze our traffic. If you allow us to do so, we also inform our social media, advertising and analysis partners about your use of our website, You can decise for yourself which categories you you want to deny or allow. Please note that based on your settings not all functionalities of the site are available. View our privacy policy.

Medical Science Monitor eISSN: 1643-3750
Medical Science Monitor eISSN: 1643-3750