Background
One potential complication of interference devices to fix anterior cruciate ligament (ACL) grafts is divergence between the implant and the graft-tunnel. Tunnel-screw divergence >15-30 degrees significantly reduces graft-tunnel fixation and may compromise the stability of the knee.
Material and Method
A complication of the IntraFix device is presented. A 43-year-old male sustained a primary ACL reconstruction. The autologous hamstring graft was fixed in the tibia with the IntraFix device. The procedure went without apparent incidents or complications. During postoperative follow-up he developed pain and crepitus over the anteromedial aspect of the operated knee. A flexion contracture slowly developed up to 10 degrees and did not improve with aggressive physiotherapy. At eight postoperative weeks, second-look arthroscopy was performed. The intraoperative findings were a healthy-looking ACL graft and a grade III chondral injury in the medial femoral condyle (MFC) caused by the tip of the IntraFix device, which protruded through the medial tibial plateau. During flexion-extension motion of the knee, the tip eroded the chondral surface of the MFC. This was treated with mechanical and thermal chondroplasty and the tip of the IntraFix device was burred flat to match the surface of the medial tibial plateau. The patient's symptoms improved.
Results
Conclusions
Divergence of the interference screws in the femur is not uncommon, but on the tibia it is rare. Although this technical error did not affect stability, it did produce intrarticular injury. Strict adherence to the surgical technique is recommended and several hints on how to avoid complications with this implant are suggested.
Keywords: Tibia - surgery, Tissue Transplantation - methods, Tendons - transplantation, Intraoperative Complications - surgery, Cartilage - injuries, Bone Screws - adverse effects, Anterior Cruciate Ligament - surgery
Editorial
01 March 2024 : Editorial
Editorial: First Regulatory Approvals for CRISPR-Cas9 Therapeutic Gene Editing for Sickle Cell Disease and Transfusion-Dependent β-ThalassemiaDOI: 10.12659/MSM.944204
Med Sci Monit 2024; 30:e944204
In Press
18 Mar 2024 : Clinical Research
Sexual Dysfunction in Women After Tibial Fracture: A Retrospective Comparative StudyMed Sci Monit In Press; DOI: 10.12659/MSM.944136
21 Feb 2024 : Clinical Research
Potential Value of HSP90α in Prognosis of Triple-Negative Breast CancerMed Sci Monit In Press; DOI: 10.12659/MSM.943049
22 Feb 2024 : Review article
Differentiation of Native Vertebral Osteomyelitis: A Comprehensive Review of Imaging Techniques and Future ...Med Sci Monit In Press; DOI: 10.12659/MSM.943168
23 Feb 2024 : Clinical Research
A Study of 60 Patients with Low Back Pain to Compare Outcomes Following Magnetotherapy, Ultrasound, Laser, ...Med Sci Monit In Press; DOI: 10.12659/MSM.943732
Most Viewed Current Articles
16 May 2023 : Clinical Research
Electrophysiological Testing for an Auditory Processing Disorder and Reading Performance in 54 School Stude...DOI :10.12659/MSM.940387
Med Sci Monit 2023; 29:e940387
17 Jan 2024 : Review article
Vaccination Guidelines for Pregnant Women: Addressing COVID-19 and the Omicron VariantDOI :10.12659/MSM.942799
Med Sci Monit 2024; 30:e942799
14 Dec 2022 : Clinical Research
Prevalence and Variability of Allergen-Specific Immunoglobulin E in Patients with Elevated Tryptase LevelsDOI :10.12659/MSM.937990
Med Sci Monit 2022; 28:e937990
01 Jan 2022 : Editorial
Editorial: Current Status of Oral Antiviral Drug Treatments for SARS-CoV-2 Infection in Non-Hospitalized Pa...DOI :10.12659/MSM.935952
Med Sci Monit 2022; 28:e935952